Z-0596-2016 Class II Terminated

Recalled by Philips Ultrasound, Inc. — Bothell, WA

FDA device recall Z-0596-2016 was initiated by Philips Ultrasound, Inc. on Dec 9, 2015 and is designated Class II. Reason for recall: Color Flow direction is displayed incorrectly in Lumify 1.0. The system displayed Color Flow direction does not correctly represent the annotated Color Bar or Velocity Markers. (e.g.) When the Color… The recall status is terminated (terminated Aug 8, 2016). Affected quantity: 21 units in the US only.

Recall details

Product Type
Devices
Report Date
Jan 13, 2016
Initiation Date
Dec 9, 2015
Termination Date
Aug 8, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21 units in the US only

Product description

Philips Lumify Diagnostic Ultrasound, Catalogue Number: 795216 Part Number: 989605449841 with 453561845331 (software version 1.0)

Reason for recall

Color Flow direction is displayed incorrectly in Lumify 1.0. The system displayed Color Flow direction does not correctly represent the annotated Color Bar or Velocity Markers. (e.g.) When the Color Bar conveys that Red is to be displayed for color flow toward the Transducer, the system displays Blue for color flow toward the Transducer. This could result in misdiagnosis in some studies.

Distribution pattern

distributed in CA, CT, ND, NV, OR, RI, TN, and WA.

Code information

System Serial numbers: USD15L0177; USD15L0178; USD15L0179; USD15L0297; USD15L0298; USN15L0789; USN15L0862; USN15L0863; USN15L0864; USN15L0865; USN15L0866; USN15L0867; USN15L0917; USN15L0918; USN15L0988; USO15L0563; USO15L0564; USO15L0565; USO15L0566; USO15L1087; and USO15L1088.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0596-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Ultrasound, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Ultrasound, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Ultrasound, Inc. →

Source: openFDA record for Z-0596-2016. Report a data issue: hello@regdatalab.com.