Z-0596-2019 Class II Terminated

Recalled by Angiodynamics Inc. (Navilyst Medical Inc.) — Glens Falls, NY

Recall Details

Product Type
Devices
Report Date
December 19, 2018
Initiation Date
October 16, 2018
Termination Date
June 26, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Box of 4 units

Product Description

BioFlo PICC with ENDEXO and PASV, ( RS 5F DL BIOFLO PASV - MEMORIAL HOSPITAL SYSTEM (LTP) PG) Cat. no. 60M183481

Reason for Recall

specific lots of Xcela Power Injectable and BioFlo PICCs are labelled with incorrect expiration dates. The labelled expiration date extends the shelf life of the product beyond the date supported by validation testing. Although the product under recall has not yet reached its validated expiry date, removal of the affected product is warranted to prevent use beyond the validated expiry date.

Distribution Pattern

Distributed to accounts in Florida. Foreign distribution to Great Britain.

Code Information

5231609