Z-0599-2014 Class II Terminated
FDA device recall Z-0599-2014 was initiated by Asante Solutions, Inc. on Nov 19, 2013 and is designated Class II. Reason for recall: Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended. The recall status is terminated (terminated Apr 24, 2014). Affected quantity: 1979 kits.
Recall details
- Product Type
- Devices
- Report Date
- Jan 8, 2014
- Initiation Date
- Nov 19, 2013
- Termination Date
- Apr 24, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1979 kits
Product description
Reason for recall
Distribution pattern
Nationwide Distribution.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0599-2014
- openFDA file date
- Nov 19, 2026