Z-0599-2018 Class II Terminated

Recalled by 3M Company - Health Care Business — Saint Paul, MN

Recall Details

Product Type
Devices
Report Date
February 21, 2018
Initiation Date
January 23, 2018
Termination Date
August 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
36 cases of 5 units

Product Description

3M Bair Hugger(M) Normothermia System, Sterile Full Access Underbody, Model 63700 The 3M Bair Hugger Temperature Management System is intended to prevent and treat hypothermia. In addition, the temperature management system can be used to provide patient thermal comfort when conditions exist that may cause patients to become too warm or too cold. The temperature management system can be used with adult and pediatric patients.

Reason for Recall

Adesign change to 3M(TM) Bair Hugger(TM) Full Access Underbody Warming Blanket, Models 63500 and 63700, can make these blankets more susceptible to unintended occlusion of airflow resulting in limited or partial inflation during use.

Distribution Pattern

Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, Chile, Costa Rica, Deutschland GMBH, Dominican, Ecuador, Hong Kong, Indonesia, Japan, Korea, Paraguay, Philippines, New Zealand, Shanghai, Thailand, and Uruguay.

Code Information

Lot Numbers: FR10361