Z-2640-2023 Class II Ongoing

Recalled by 3M Company - Health Care Business — Saint Paul, MN

FDA device recall Z-2640-2023 was initiated by 3M Company - Health Care Business on Aug 22, 2023 and is designated Class II. Reason for recall: Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V)… The recall status is ongoing. Affected quantity: 77,280 devices.

Recall details

Product Type
Devices
Report Date
Oct 4, 2023
Initiation Date
Aug 22, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
77,280 devices

Product description

3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air removal steam sterilization processes

Reason for recall

Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.

Code information

UDI/DI: 50707387787283( (shipper); Lot Codes: 41482VF, exp. 33P4ND; 33P4YN, exp. 01/02/2025

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2640-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for 3M Company - Health Care Business? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for 3M Company - Health Care Business: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about 3M Company - Health Care Business →

Source: openFDA record for Z-2640-2023. Report a data issue: hello@regdatalab.com.