Z-1169-2024 Class II Ongoing

Recalled by 3M Company - Health Care Business — Saint Paul, MN

FDA device recall Z-1169-2024 was initiated by 3M Company - Health Care Business on Jan 12, 2024 and is designated Class II. Reason for recall: 3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured… The recall status is ongoing. Affected quantity: 52,833,750 blades.

Recall details

Product Type
Devices
Report Date
Feb 28, 2024
Initiation Date
Jan 12, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
52,833,750 blades

Product description

3M Specialty Blade Assembly, REF 9660,70-2011-8462-2, 7100213114, used with used with 3M Surgical Clipper Catalog Number 9661L

Reason for recall

3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41¿C (105.8¿F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes without applying to skin, the Model 9690 blade maximum temperature may reach 60¿C (140¿F) and the Model 9660 blade maximum temperature may reach 77¿C (170.6¿F)."

Distribution pattern

Worldwide distribution.

Code information

UDI/DI 50707387789003 (shipper), 30707387789009 (tray), all lot codes

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1169-2024
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for 3M Company - Health Care Business? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for 3M Company - Health Care Business: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about 3M Company - Health Care Business →

Source: openFDA record for Z-1169-2024. Report a data issue: hello@regdatalab.com.