Z-0612-2017 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc — East Walpole, MA

FDA device recall Z-0612-2017 was initiated by Siemens Healthcare Diagnostics, Inc on Oct 19, 2016 and is designated Class II. Reason for recall: CA 19-9 assay on the ADVIA Centaur CP does not meet the correlation to the CA 19-9 assay on the ADVIA Centaur/XP/XPT systems as stated in the Instructions for Use. The recall status is terminated (terminated Mar 2, 2023). Affected quantity: 18,733 kits.

Recall details

Product Type
Devices
Report Date
Nov 23, 2016
Initiation Date
Oct 19, 2016
Termination Date
Mar 2, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18,733 kits

Product description

ADVIA Centaur CA 19-9 Assay (50 test kit)-for use, with the ADVIA Centaur and ADVIA Centaur XP, ADVIA Centaur XPT and ADVIA Centaur CP systems. Siemens Material Number(SMN): 10491379 In the 510(k) database the CA 19-9 assay is not CLIA categorized on the ADVIA Centaur XPT instrument. Siemens stated that the assay is being marketed on the XPT following FDAs Reagent Replacement Policy. Siemens plans to request CLIA Categorization for the CA19-9 assay on XPT the week of 11/21/2016.

Reason for recall

CA 19-9 assay on the ADVIA Centaur CP does not meet the correlation to the CA 19-9 assay on the ADVIA Centaur/XP/XPT systems as stated in the Instructions for Use.

Distribution pattern

Nationwide Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Germany, Great Britain, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Latvia, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mexico, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, Vatican City, Vietnam

Code information

Kit lots ending in: 380 Exp Date 2016-10-21; 382/ 2017-01-09; 384 2017-02-06; 386 2017-03-24 ; 388 2017-05-14 ; 390 2017-06-12 . And all future lots until the issue is resolved. Siemens will also include a notecard in future lots of reagents referring customers to the communications .

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0612-2017
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Healthcare Diagnostics, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Healthcare Diagnostics, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Healthcare Diagnostics, Inc →

Source: openFDA record for Z-0612-2017. Report a data issue: hello@regdatalab.com.