Z-0613-2020 Class II Terminated

Recalled by Integra LifeSciences Corp. — Plainsboro, NJ

Recall Details

Product Type
Devices
Report Date
December 18, 2019
Initiation Date
October 21, 2019
Termination Date
May 7, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
820

Product Description

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Reason for Recall

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Distribution Pattern

US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK

Code Information

US Ops Manual P/N 60905769