Recalled by Atrium Medical Corporation — Merrimack, NH

Recall Details

Product Type
Devices
Report Date
January 3, 2024
Initiation Date
November 1, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
835 cases total (6 units/case)

Product Description

Atrium Oasis Drain Single W / AC, Part No. 3600-100

Reason for Recall

A sterilization nonconformance prematurely aged the product by 48.3 days, resulting in the expiration date on the label not being accurate. This nonconforming product was inadvertently distributed to customers. Increased risk to patients and healthcare providers using the product after the shortened expiration date of November 17, 2025.

Distribution Pattern

Domestic distribution nationwide. No foreign distribution.

Code Information

UDI-DI 00650862110012 Lots 492065 and 492075