Recalled by Integra LifeSciences Corp. — Plainsboro, NJ

FDA device recall Z-0622-2014 was initiated by Integra LifeSciences Corp. on Nov 13, 2013 and is designated Class II. Reason for recall: Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type packages were found prior to use but after the… The recall status is terminated (terminated Feb 13, 2015). Affected quantity: 78 packages.

Recall details

Product Type
Devices
Report Date
Jan 15, 2014
Initiation Date
Nov 13, 2013
Termination Date
Feb 13, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
78 packages

Product description

Integra Large Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Made in the USA NL 850-1132 Provides access to the lateral cerebral ventricles via hypodermic puncture. It is useful in obtaining CSF (cerebral spinal fluid_ samples for cytological and chemical studies, for monitoring ventricular fluid pressure and for ventricular drainage. The reservoir provides easy access to the lateral ventricles and to cystic tumors for the injections of chemotherapeutic agents and/or radio-isotopes. The rickham reservoir may be utilized in hydrocephalic patients as a component in systems designed to shunt CSF from the lateral ventricles into either the right atrium of the heart of the peritoneum.

Reason for recall

Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type packages were found prior to use but after the sterile package was open.

Distribution pattern

Worldwide Distribution - USA including FL, NE, WI, WA, CT, ND, AL, NY, NH, OR, OK, MN, and TX. Internationally to Belgium, Czech Republic, Germany, France, Great Britian, Italy, Poland, and Portugal.

Code information

1094085, 1101479, 1101647, 1102403, 1102807, 1104174, 1110043, 1120817, 1130797, 1131331, 1131660

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0622-2014
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0622-2014. Report a data issue: hello@regdatalab.com.