Z-0629-2019 Class II Terminated

Recalled by Jarvik Heart Inc — New York, NY

Recall Details

Product Type
Devices
Report Date
December 26, 2018
Initiation Date
October 15, 2018
Termination Date
May 1, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
274

Product Description

Jarvik 2000 Ventricular Assist System external components: JHI-212 Y Cable, JHI-202 Lithium-ion Battery Cable, JHI-204 Extensions Cable, JHI-201 Pedestal Cable

Reason for Recall

There is a potential for the male connectors on external cables to become damaged. This could include a detached male connector housing, misaligned connector, cracked connector housing, and bent pins.

Distribution Pattern

Products were distributed to the following US states: FL, NY, and UT. Products were distributed to the following foreign countries: Croatia, Czech Republic, France, Hungary, India, Italy, and Japan.

Code Information

All systems/cables