Z-0630-2026 Class II Ongoing

Recalled by B Braun Medical Inc — Bethlehem, PA

FDA device recall Z-0630-2026 was initiated by B Braun Medical Inc on Oct 29, 2025 and is designated Class II. Reason for recall: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). The recall status is ongoing. Affected quantity: 43,452 units.

Recall Details

Product Type
Devices
Report Date
Dec 3, 2025
Initiation Date
Oct 29, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
43,452 units

Product Description

Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. EXTENSION SET, 18 IN; Catalog Number: 490448. 2. EXTENSION SET, 52 IN.; Catalog Number: 490465. 3. EXTENSION SET 49 IN.; Catalog Number: 490486. 4. EXTENSION SET, 12 CARESTITE LADS; Catalog Number: 490577.

Reason for Recall

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Distribution Pattern

Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.

Code Information

1. Catalog Number: 490448; Primary UDI-DI: 04046964848858; Unit of Dose UDI-DI: 04046964848841. Shelf Life: Earliest Exp of Comp or 36mths. 2. Catalog Number: 490465; Primary UDI-DI: 04046964950735; Unit of Dose UDI-DI: 04046964950728. Shelf Life: Earliest Exp of Comp or 36mths. 3. Catalog Number: 490486; Primary UDI-DI: 04046964991257; Unit of Dose UDI-DI: 04046964991240. Shelf Life: Earliest Exp of Comp or 24mths. 4. Catalog Number: 490577; Primary UDI-DI: 04046955287536; Unit of Dose UDI-DI: 04046955287529. Shelf Life: Earliest Exp of Comp or 36mths.