Z-0635-2013 Class III Terminated
FDA device recall Z-0635-2013 was initiated by Siemens Healthcare Diagnostics Inc on Nov 21, 2012 and is designated Class III. Reason for recall: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F). The recall status is terminated (terminated Nov 6, 2014). Affected quantity: 7865.
Recall details
- Product Type
- Devices
- Report Date
- Jan 16, 2013
- Initiation Date
- Nov 21, 2012
- Termination Date
- Nov 6, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7865
Product description
Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry Catalog Number:1360 Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX¿ positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes
Reason for recall
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).
Distribution pattern
Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,
Code information
Carton/Bottle Lot Number: C9093061B P9093061B C9094081B P9094081B C0097091B P0097091B C0100101C P0100101C C0103042F P0103042F C0106062B P0106062B C91 01022D P91 01022D CP9105052B P9105052B C9108092A P9108092A