Z-0636-2020 Class II Terminated

Recalled by Integra LifeSciences Corp. — Plainsboro, NJ

FDA device recall Z-0636-2020 was initiated by Integra LifeSciences Corp. on Oct 21, 2019 and is designated Class II. Reason for recall: The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for… The recall status is terminated (terminated May 7, 2020). Affected quantity: 820.

Recall details

Product Type
Devices
Report Date
Dec 18, 2019
Initiation Date
Oct 21, 2019
Termination Date
May 7, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
820

Product description

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Reason for recall

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Distribution pattern

US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK

Code information

Portuguese OPS Manual P/N 60905207

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0636-2020
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0636-2020. Report a data issue: hello@regdatalab.com.