Z-0638-2013 Class III Terminated
FDA device recall Z-0638-2013 was initiated by Siemens Healthcare Diagnostics Inc on Nov 21, 2012 and is designated Class III. Reason for recall: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F). The recall status is terminated (terminated Nov 6, 2014). Affected quantity: 700.
Recall details
- Product Type
- Devices
- Report Date
- Jan 16, 2013
- Initiation Date
- Nov 21, 2012
- Termination Date
- Nov 6, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 700
Product description
Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry (Distributed OUS Only) Catalog Number: 1360L Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX¿ positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes
Reason for recall
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).
Distribution pattern
Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,
Code information
Carton/Bottle Lot Number: C9094081C P9094081C C0100101A P0100101A C9101022C P9101022C
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0638-2013
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026