Z-0640-2013 Class II Terminated
FDA device recall Z-0640-2013 was initiated by Siemens Healthcare Diagnostics Inc on Nov 21, 2012 and is designated Class II. Reason for recall: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F). The recall status is terminated (terminated Nov 6, 2014). Affected quantity: 1130.
Recall details
- Product Type
- Devices
- Report Date
- Jan 16, 2013
- Initiation Date
- Nov 21, 2012
- Termination Date
- Nov 6, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1130
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,