Z-0642-2024 Class I Ongoing

Recalled by Globus Medical, Inc. — Audubon, PA

FDA device recall Z-0642-2024 was initiated by Globus Medical, Inc. on Dec 7, 2023 and is designated Class I. Reason for recall: A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement. The recall status is ongoing. Affected quantity: 9.

Recall details

Product Type
Devices
Report Date
Jan 24, 2024
Initiation Date
Dec 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9

Product description

ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100

Reason for recall

A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.

Distribution pattern

Domestic distribution to the following states: CA, CO, MT, SC, WA. International distribution to Germany and UK.

Code information

UDI-DI 00193982465155 Serial Numbers FPFF-0101 FPFF-0102 FPFF-0103 FPFF-0104 FPFF-0105 FPFF-0106 FPFF-0107 FPFF-0108 FPFF-0109

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0642-2024
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Globus Medical, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0642-2024. Report a data issue: hello@regdatalab.com.