Z-0647-2013 Class III Terminated

Recalled by Siemens Healthcare Diagnostics — Tarrytown, NY

FDA device recall Z-0647-2013 was initiated by Siemens Healthcare Diagnostics on Nov 28, 2012 and is designated Class III. Reason for recall: The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C)… The recall status is terminated (terminated Oct 28, 2013). Affected quantity: Lot 831=19 units (all foreign consignees) and Lot 832=2 units (all foreign consignees).

Recall details

Product Type
Devices
Report Date
Jan 16, 2013
Initiation Date
Nov 28, 2012
Termination Date
Oct 28, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Lot 831=19 units (all foreign consignees) and Lot 832=2 units (all foreign consignees)

Product description

Immunoradiometric assay.

Reason for recall

The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 (level G) lot 035. The affected calibrator vials can be identified by a darker yellow color than normal. The TSI3 (level A), TSI7 (level E), and TSIX (level H)

Distribution pattern

Worldwide Distribution - USA including CA, GA, IL, IN, MI, MN, NY, and Puerto Rico. Internationally to Agentina, Bolivia, Brazil, Canada, Chile, Guatemala, South Korea, Philippines, Paraguay, Taiwan, and Uraguay,

Code information

Lots 831 and 832

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0647-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0647-2013. Report a data issue: hello@regdatalab.com.