Z-0648-2020 Class II Terminated

Recalled by Integra LifeSciences Corp. — Plainsboro, NJ

Recall Details

Product Type
Devices
Report Date
December 18, 2019
Initiation Date
October 21, 2019
Termination Date
May 7, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Reason for Recall

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Distribution Pattern

US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK

Code Information

Polish OPS Manual P/N 60905213