Z-0651-2014 Class II Terminated

Recalled by Midmark Corp Dba Progeny Inc — Lincolnshire, IL

FDA device recall Z-0651-2014 was initiated by Midmark Corp Dba Progeny Inc on Sep 20, 2013 and is designated Class II. Reason for recall: An error was identified in the Real Time Controller (RTC) firmware versions 3.0, 3.1, 3.2, and 3.3 used in Vantage Panoramic X-Ray systems. The error may cause the column of the Vantage Panoramic… The recall status is terminated (terminated Dec 27, 2017). Affected quantity: 58 units.

Recall details

Product Type
Devices
Report Date
Jan 15, 2014
Initiation Date
Sep 20, 2013
Termination Date
Dec 27, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
58 units

Product description

The Vantage Digital Panoramic System delivers high-value features standard on every system: - Five pre-programmed exam settings, including bitewing, minimize exam set-up time and allow routine diagnostic exams to be performed extraorally for improved clinical efficiency and patient experience. - Progeny Clarity Enhanced Filter Set provides superior diagnostic imaging quality - Exclusive Focal Trough Alignment reminder assists in consistent patient positioning - Progeny Imaging software included for download on an unlimited number of operatory workstations with no additional costs - Exclusive VantageTrust remote support and training service offered at no charge for the first 60 days of use - Ceph upgradeable for expanded extraoral exam capabilities - The Progeny Vantage Panoramic System s proudly designed and manufactured in the USA; 2013 Midmark Corporation; 675 Heathrow Drive Lincolnshire, IL 60069; 847-415-9800 Toll-free 888-924-3800

Reason for recall

An error was identified in the Real Time Controller (RTC) firmware versions 3.0, 3.1, 3.2, and 3.3 used in Vantage Panoramic X-Ray systems. The error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. failure to stop").

Distribution pattern

Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, CA, CO, CT, FL, GA, IL, KS, MA, NC, NH, NJ, OH, OK, PA, SC, TN, TX, VA, and WI, and the countries of Australia, Canada, and Russia.

Code information

Vantage Panoramic X-Ray System firmware versions: 3.0, 3.1, 3.2, and 3.3. Vantage Panoramic X-Ray System Catalog Numbers and Serial Numbers: Catalog Number V5000, Serial Numbers ZM81025, ZM81026, ZM81030, ZZ08005, ZZ08006, ZZ08007, ZZ08008, ZZ08009, ZZ08010, ZZ08011, ZZ08012, ZZ08013, ZZ08015, ZZ08016, ZZ08017, Z08018, ZZ08019, ZZ08020, ZZ08021, ZZ08022, ZZ08023, ZB51001, ZB51002, ZB51006, ZB51007, ZT18018, ZY49014, ZK58019, ZS54011, ZP01043, ZV78016, ZE09006, ZN44018, ZS59007, ZS59001, ZW15007, ZW34003, ZW34012, ZW34018, ZY49012, ZY49015, ZZ02010, ZM81022, ZS54012, ZE09019, ZM81017, ZT18004, ZT18015, ZN44020, ZE09012, ZM81023, ZT27012; Catalog Number V5050, Serial Number ZZ08014; Catalog Number V5100, Serial Numbers ZM81027, ZM81028, ZZ08003; Catalog Number V5150, Serial Number ZZ08002; Catalog Number V5150C, Serial Number ZM81016

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0651-2014
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Midmark Corp Dba Progeny Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0651-2014. Report a data issue: hello@regdatalab.com.