Z-0656-2020 Class II Terminated

Recalled by Ethicon Endo-Surgery Inc — Blue Ash, OH

Recall Details

Product Type
Devices
Report Date
December 18, 2019
Initiation Date
November 7, 2019
Termination Date
April 23, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
48,428

Product Description

Ligamax 5 mm Endoscopic Multiple Clip Applier

Reason for Recall

A potential exists that certain LIGAMAX devices within the impacted lots may have small holes in the Tyvek lidding which could result in a breach of sterility.

Distribution Pattern

US: Nationwide International: ARGENTINA ARUBA AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CHINA COLOMBIA DENMARK ESTONIA FINLAND FRANCE GERMANY INDIA INDONESIA IRELAND JAPAN KOREA KUWAIT MALAYSIA MEXICO NETHERLANDS NEW CALEDONIA NEW ZEALAND PANAMA PHILIPPINES PUERTO RICO QATAR SAUDI ARABIA SINGAPORE SLOVENIA SWEDEN SWITZERLAND TAIWAN UK

Code Information

Product Code: 11EL5ML Lot Numbers: T93X4R, T93X9D, T93Y2Z, T93Y70, T93Y96, T93Z52, T93Z76, T9401C, T9408F, T94118, T9416M, T9422K, T9426Z, T94402, T94536, T94567, T9461M, T94667, T94721, T94758, T94807, T9484Y, T9487J, T9491A T94A9K, T94C2Z, T94C8N, T94D2N, T94E66, T94F3C, T94F7F, T94G99 T94H1N, T94H78