Z-2356-2021 Class II Ongoing

Recalled by Ethicon Endo-Surgery Inc — Blue Ash, OH

FDA device recall Z-2356-2021 was initiated by Ethicon Endo-Surgery Inc on Jul 20, 2021 and is designated Class II. Reason for recall: An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns The recall status is ongoing. Affected quantity: 68960 units.

Recall details

Product Type
Devices
Report Date
Sep 1, 2021
Initiation Date
Jul 20, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
68960 units

Product description

HARMONICHD 1000i Shears (36cm Shaft Length)-indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. Product Code: HARHD36

Reason for recall

An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns

Distribution pattern

Worldwide distribution - US Nationwide distribution and the countries of Austria, Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, United Arab Emirates.

Code information

Lot Codes: U94Y9V U95C2A U94Z98 U95427 U9516E U95599 U9503G U95856 U94Y9W U95L0A U9518D U95526 U9507E U9548R U95127 U95525 U9530P U95T2Z U95366 U95R4F U9521Z U95F2T U9554U U95F8W U9564G U95U6A U95524 U95E6F U9512D U95K4D U94Z49 U95F07 U95A9X U95F6N U9569W U95E1U U9571P U95E72 U9587D U95E29 U95815 U95T90 U95814 U95Z67 U95754 U95Y7G U95D4T U9523L U95566 U9530X U9543R GTIN: 10705036015055

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2356-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ethicon Endo-Surgery Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Ethicon Endo-Surgery Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Ethicon Endo-Surgery Inc →

Source: openFDA record for Z-2356-2021. Report a data issue: hello@regdatalab.com.