Z-0202-2020 Class I Terminated

Recalled by Ethicon Endo-Surgery Inc — Blue Ash, OH

FDA device recall Z-0202-2020 was initiated by Ethicon Endo-Surgery Inc on Oct 3, 2019 and is designated Class I. Reason for recall: The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple… The recall status is terminated (terminated Jun 3, 2021). Affected quantity: 5040 units.

Recall details

Product Type
Devices
Report Date
Nov 6, 2019
Initiation Date
Oct 3, 2019
Termination Date
Jun 3, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5040 units

Product description

ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 440mm shaft, Product Code PLEE60A

Reason for recall

The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

Distribution pattern

Worldwide distribution. US Nationwide. Belgium, Brazil, Canada, Chile, Colombia, Japan, Kuwait, Saudi Arabia, and Singapore.

Code information

Lot: T93Z1G, T93Y4M, T94045, T94117, T93X17, T94087, T93Z2X, T9405L, T94253, T93X95, T93Z75, T93Z2W, T9413Z

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0202-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ethicon Endo-Surgery Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Ethicon Endo-Surgery Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Ethicon Endo-Surgery Inc →

Source: openFDA record for Z-0202-2020. Report a data issue: hello@regdatalab.com.