Z-0203-2020 Class I Terminated

Recalled by Ethicon Endo-Surgery Inc — Blue Ash, OH

Recall Details

Product Type
Devices
Report Date
November 6, 2019
Initiation Date
October 3, 2019
Termination Date
June 3, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2112 units

Product Description

ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 340mm shaft, Product Code PSEE60A

Reason for Recall

The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

Distribution Pattern

Worldwide distribution. US Nationwide. Belgium, Brazil, Canada, Chile, Colombia, Japan, Kuwait, Saudi Arabia, and Singapore.

Code Information

Lot: T93Z5W, T93Z5X, T9405V, T9405W, T93Z3F, T9401L, T93Y8X, T94008, T9400D, T93Z5R