Z-0667-2020 Class II Terminated
FDA device recall Z-0667-2020 was initiated by Boston Scientific Corporation on Nov 18, 2019 and is designated Class II. Reason for recall: During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from… The recall status is terminated (terminated Jun 25, 2020). Affected quantity: 139.
Recall details
- Product Type
- Devices
- Report Date
- Dec 18, 2019
- Initiation Date
- Nov 18, 2019
- Termination Date
- Jun 25, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 139
Product description
Reason for recall
Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, CO, CT, FL, NJ, NY, MA, MD, MN, TN, TX and countries of Canada, India, Australia, Great Britain, Greece, Spain.