Z-0668-2020 Class II Terminated

Recalled by Boston Scientific Corporation — Marlborough, MA

FDA device recall Z-0668-2020 was initiated by Boston Scientific Corporation on Nov 18, 2019 and is designated Class II. Reason for recall: During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from… The recall status is terminated (terminated Jun 25, 2020). Affected quantity: 179.

Recall details

Product Type
Devices
Report Date
Dec 18, 2019
Initiation Date
Nov 18, 2019
Termination Date
Jun 25, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
179

Product description

Habib EUS RFA, 20mm electrode active tip, Material Number/UPN: 6700 - Product Usage: The Habib EUS RFA is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radiofrequency generator.

Reason for recall

During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from the Habib EUS RFA active tip to the introducer device. This capacitive coupling could result in thermal injury to tissue outside of the targeted treatment area due to its use with an un-insulated EUS-FNA needle requiring little or no medical intervention.

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, CO, CT, FL, NJ, NY, MA, MD, MN, TN, TX and countries of Canada, India, Australia, Great Britain, Greece, Spain.

Code information

Material Number/UPN: 6700 GTIN: 5060271140596 Batches: 17-C01 (Expiration: 10-Jul-2020), 17-C03 (Expiration: 12-Nov-2020), 17-C04 (Expiration: 18-Feb-2021), 17-C05 (Expiration: 18-Feb-2021), 17-C09 (Expiration: 19-Feb-2021)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0668-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Boston Scientific Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Boston Scientific Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Boston Scientific Corporation →

Source: openFDA record for Z-0668-2020. Report a data issue: hello@regdatalab.com.