Z-0673-2017 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-0673-2017 was initiated by Roche Diagnostics Operations, Inc. on Oct 31, 2016 and is designated Class II. Reason for recall: Incorrect Urea results may be generated sporadically on cobas b 221<6>Roche OMNI S6 system and cannot be detected by QC. QC measurement results can be affected as well. This issue is only detectable… The recall status is terminated (terminated Sep 18, 2017). Affected quantity: 97.

Recall Details

Product Type
Devices
Report Date
Dec 7, 2016
Initiation Date
Oct 31, 2016
Termination Date
Sep 18, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
97

Product Description

MSS CASSETTE GLU/LAC/UREA (BUN) as a part of the Roche OMNI S Analyzer: cobas b 221<6>Roche OMNI S6 system The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate

Reason for Recall

Incorrect Urea results may be generated sporadically on cobas b 221<6>Roche OMNI S6 system and cannot be detected by QC. QC measurement results can be affected as well. This issue is only detectable if comparison result of a second measurement is available. The cause for this issue is an error in the software measurement procedure. In children, decreased urea results can point at inborn disorders or urea cycle. Therefore, false normal urea values could lead to delayed/incorrect diagnosis of such a disorder. Besides the patients most at risk, the risk for the overall patient population with falsely low or high values is considered remote. No severe adverse health effects are expected; but, unnecessary diagnostic measures may be initiated. Falsely normal results may result in further necessary examination not being carried out, with the possibility of a delay in the diagnosis and treatment. A medical risk cannot entirely be excluded.

Distribution Pattern

Nationwide Distribution including ALCA CO CT FL IL IN KS MD MI MO NC NE NJ NM NY OH OK PA PR SC TX VA and WI

Code Information

Lot Number(s): All lots within shelf life. Expiration Date(s): All

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0673-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.