Z-0682-2017 Class II Terminated

Recalled by Sorin Group USA, Inc. — Arvada, CO

FDA device recall Z-0682-2017 was initiated by Sorin Group USA, Inc. on Aug 1, 2014 and is designated Class II. Reason for recall: On August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use… The recall status is terminated (terminated Apr 20, 2017). Affected quantity: 1,618 units in the US; 129 units outside of US.

Recall details

Product Type
Devices
Report Date
Dec 7, 2016
Initiation Date
Aug 1, 2014
Termination Date
Apr 20, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,618 units in the US; 129 units outside of US

Product description

ST¿CKERT HeaterCooler System 3T; Item Number 160281, HeaterCooler 3T, 240V/60Hz; Item Number 160282, HeaterCooler 3T, 208V/60Hz; Item Number 160285, HeaterCooler 3T, 120V/ 60Hz The Heater Cooler 3T is used with a St¿ckert S3 heart lung machine and / or any other heart lung machine featuring a separate temperature control for extracorporeal perfusion of durations up to 6 hours

Reason for recall

On August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU).

Distribution pattern

Worldwide Distribution - US (Nationwide and Puerto Rico) and Internationally to Canada, Colombia, Ecuador, El Salvador, Germany, Honduras, Mexico, Saudi Arabia, Trinidad, and Tobago.

Code information

All Serial Numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0682-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0682-2017. Report a data issue: hello@regdatalab.com.