Z-0697-2021 Class II Ongoing
FDA device recall Z-0697-2021 was initiated by Haag-Streit USA Inc on Nov 20, 2020 and is designated Class II. Reason for recall: Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing… The recall status is ongoing. Affected quantity: 3 units US.
Recall details
- Product Type
- Devices
- Report Date
- Dec 30, 2020
- Initiation Date
- Nov 20, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 units US
Product description
Reason for recall
Distribution pattern
US Nationwide distribution in the state of OK.