Z-0697-2021 Class II Ongoing

Recalled by Haag-Streit USA Inc — Mason, OH

FDA device recall Z-0697-2021 was initiated by Haag-Streit USA Inc on November 20, 2020 and is designated Class II. Reason for recall: Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing … The recall status is ongoing. Affected quantity: 3 units US.

Recall Details

Product Type
Devices
Report Date
December 30, 2020
Initiation Date
November 20, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units US

Product Description

Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination with the operating microscopes HS Hi-R NEO 900 (Ref no. 657 820) and HS Hi-R NEO 900A (Ref no. 657 821), with software REF 615 588 versions 2.0 to 3.3 - Product Usage: use in ophthalmology and in other disciplines with vertical direction of view

Reason for Recall

Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing the button, result in potential risk to the patient's eye

Distribution Pattern

US Nationwide distribution in the state of OK.

Code Information

Floor stands FS 2-11, FS 2-15 with the software REF 615 588 in versions 2.0 to 3.3. FS 2-11 Ref. # 615H511 Serial Numbers:From 1001; FS 2-15 Ref. #615H515 Serial Numbers From 1001; HS Hi-R NEO 900 Ref. # 657 Serial Numbers 820 from 101; HS Hi-R NEO 900A Ref # 657 821 Serial Numbers from 101 Unique Device Identifier (UDI) HS Hi-R NEO 900 +EMWM657 820 0/$+serial no. HS Hi-R NEO 900A +EMWM657 821 0/$+serial no. FS 2 11 +EMWM 615H5110/$+serial no. FS 2-15 +EMWM 615H515 0/$+serial no.