Z-0699-2021 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 30, 2020
- Initiation Date
- August 31, 2020
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 19850
Product Description
42 IN (107 cm) Appx 5.0 ml, 20 Drop Admin Set, Spiros, REF: 20131-01; 31" (79 cm) Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros, Bag Hanger Drop-in Red Cap, 4 Units, REF: CH3330-C; 60 IN (152 cm) Appx 1.1 ml, Smallbore Set, MicroClave Clear, Spiros, REF: CH3353; 8" (20 cm) Appx 0.43 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Clamps, Spiros, REF: CH3367; 7" (18 cm) Appx 0.40 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, Spiros, 2 Clamps, REF: CH3379; 16" (41 cm) Appx 4.4 ml, Ext Set w/MicroClave, 0.2 Micron Filter, Clave, Spiros, Clamp, REF: CH3381; 31" (79 cm) Appx 3.3 ml, Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros w/Drop-in Red Cap, 2 Hanger, REF: CH3386; 40" (102 cm) Appx 7.3 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: CH3401; 60" (152 cm) Appx 0.62 ml, Smallbore Ext Set w/MicroClave, Clamp, Spiros w/Red Cap, REF: CH3405; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Clave, Spiros w/Red Cap, REF: CH3453; 73" (185 cm) Appx 2.1 ml, Smallbore Ext Set w/MicroClave Clear, Spiros, Clamp, REF: CH3485; 42" (107 cm) Appx 4.8 ml, 15 Drop Admin Set w/Spiros w/Red Cap, Bag Hanger, Drop-In Priming Cap, REF: CH3517; 8" (20 cm) Appx 0.49 ml, Smallbore Bifuse Ext Set w/MicroClave Clear, 2 Spiros, 3 Clamps, Priming Cap, REF: CH3523; 74" (188 cm) Smallbore Ext Set w/MicroClave, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3533; 74 IN (188 cm) Appx 0.92 ml, Smallbore Set, MicroClave, 1.2 Micron Fil, Spiros, Red Cap, REF: CH3538; 18" (46 cm) Appx 5.1 ml, Ext Set w/0.2 Micron Filter, Clave, Spiros, Clamp, REF: CH3556; 60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; 4-Way Clave Stopcock w/2 Spiros, REF: B4194; Spiros, Closed Male Connector w/Red Cap (10 Units), REF: Z7093; 47" (119 cm) 150 ml Burette Set w/MicroClave, 20 Drop In-Line Drip Chamber w/Ball Shut Off, Spiros, REF: Z7182
Reason for Recall
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Distribution Pattern
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Code Information
Lot: 4764635, 4887770, 4787215, 4877053, 4737148, 4850390, 4749863, 4849350, 4867377, 4741238, 4763286, 4774081, 4809085, 4877072, 4908343, 4719787, 4734320, 4741243, 4749867, 4854763, 4895575, 4849358, 4751063, 4763289, 4775529, 4724813, 4749880, 4854774, 4734335, 4757706, 4849376, 4764061, 4793547, 4749900, 4756873, 4777159, 4799805, 4902253, 4751908, 4793550, 4849389, 4896178, 4733498, 4764065, 4849391, 4719565, 4751101, 4790766, 4774117, 4736129, 4895706, 4858370