Z-0703-2017 Class II Terminated

Recalled by DePuy Orthopaedics, Inc. — Warsaw, IN

FDA device recall Z-0703-2017 was initiated by DePuy Orthopaedics, Inc. on Nov 8, 2016 and is designated Class II. Reason for recall: DePuy Orthopaedics, Inc. is voluntarily recalling one lot of the LCS Complete Knee Revision System VVC Insert Small 22.5mm, due to the insert being manufactured outside of specifications, which… The recall status is terminated (terminated Jun 5, 2017). Affected quantity: 1.

Recall Details

Product Type
Devices
Report Date
Dec 14, 2016
Initiation Date
Nov 8, 2016
Termination Date
Jun 5, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product Description

LCS Complete Knee Revision System VVC Insert Product Usage: LCS VVC Insert is used in Knee Revision Surgery

Reason for Recall

DePuy Orthopaedics, Inc. is voluntarily recalling one lot of the LCS Complete Knee Revision System VVC Insert Small 22.5mm, due to the insert being manufactured outside of specifications, which prevents the insert from properly seating in the mating tibial tray.

Distribution Pattern

US Distribution in the state of TN

Code Information

LCS COMP VVC INS SM 22.5MM, Product Code: 129424122, Lot# C08294, GTIN: (01)10603295024484, Expiry: 31-Mar-2021

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0703-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.