Z-0708-2021 Class II Ongoing

Recalled by ICU Medical, Inc. — San Clemente, CA

Recall Details

Product Type
Devices
Report Date
December 30, 2020
Initiation Date
August 31, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4790

Product Description

76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF: 011-CH3261; 44 cm(17 IN) Appx 2.9 ml,PUR YELLOW TRANS SET,ChemoClave,CK VLV,Spiros,PURPLE CAP, REF: 011-CH3568; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext w/2 ChemoClave, Red Ring, BCV; 297cm Ext w/ChemoClave, REF: 011-CH3600; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext, w/2 ChemoClave, Red Ring, BCV; 597 cm Ext w/ChemoClave, REF: 011-CH3601; Kit, 2 5 Gang Stopcocks w/5 MicroClave Clear, Spiros; 2 Smallbore Bifuse Ext Sets, 2 MicroClave Clear, Red Ring, Check Valve, RMLL; 2 Ext Sets, MicroClave Clear Clipped Together, RMLL, REF: 011-CH3610; Admin Set w/Clave, Spiros, REF: 011-CH3772; Admin Set w/Clave, Spiros, REF: 011-CH3773; Admin Set w/Clave, Spiros, REF: 011-CH3967; 76" (193 cm) 50 ml, Diluent Set for Channel 2 w/20 Drop Drip Chamber, Spiros, Drop-In Red Cap, REF: 011-CH4011

Reason for Recall

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Distribution Pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

Code Information

Lots: 4784390, 4784404, 4865931, 4865006, 4590283, 4774210, 4865952, 4711264, 4605516, 4723103, 4882016, 4806569 ,