Z-0713-2015 Class II Terminated

Recalled by Stryker Endoscopy — San Jose, CA

FDA device recall Z-0713-2015 was initiated by Stryker Endoscopy on Nov 17, 2014 and is designated Class II. Reason for recall: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle… The recall status is terminated (terminated Jul 17, 2015). Affected quantity: 180,573 units total.

Recall details

Product Type
Devices
Report Date
Dec 24, 2014
Initiation Date
Nov 17, 2014
Termination Date
Jul 17, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
180,573 units total

Product description

PKG, 5MM SHAFT, NEEDLE HOLDER, STRAIGHT, 45CM, P/N 0250080588 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason for recall

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Distribution pattern

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Code information

342649 343202 543358 644784 744458 747492 747493 840089 842490 847795 940117 943497 943730 943858 945716 1041182 1042740 1042741 1043533 1045231 1046109 1140858 1141657 1142152 1143125 1241809 1242140 1243099 1244074 1245022 1246067 1340117 1341249 1346653 1440696 1442806

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0713-2015
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Stryker Endoscopy? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0713-2015. Report a data issue: hello@regdatalab.com.