Z-0716-2025 Class I Ongoing

Recalled by Trokamed GmbH — Geisingen

FDA device recall Z-0716-2025 was initiated by Trokamed GmbH on Nov 27, 2024 and is designated Class I. Reason for recall: If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure. The recall status is ongoing. Affected quantity: 102 units.

Recall details

Product Type
Devices
Report Date
Jan 22, 2025
Initiation Date
Nov 27, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
102 units

Product description

Sheath, 18 Fr., 160 mm (REF: WA2PS18L), gastroenterology-urology endoscopic access overtube.

Reason for recall

If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure.

Distribution pattern

US (Domestic) distribution in state of PA and OUS (foreign) to country of Germany. .

Code information

REF: WA2PS18L; UDI-DI: 04251303810872; Lot Numbers: 154214

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0716-2025
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Trokamed GmbH? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Trokamed GmbH: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Trokamed GmbH →

Source: openFDA record for Z-0716-2025. Report a data issue: hello@regdatalab.com.