Z-0719-2021 Class II Terminated

Recalled by B Braun Medical, Inc. — Carrollton, TX

Recall Details

Product Type
Devices
Report Date
December 30, 2020
Initiation Date
November 23, 2020
Termination Date
March 14, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 units

Product Description

APEX Compounding System Control Panel Module, REF: 601238, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.

Reason for Recall

There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the ordered volume for an ingredient. This could result in a a compounded drug that does not meet the specified accuracy range for an individual ingredient.

Distribution Pattern

US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, IL, LA, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, PA, TN,TX, WA.

Code Information

All APEX Compounding System with Software Version 1.0.5.57 or earlier