Z-0728-2020 Class II Ongoing
Recalled by Cardinal Health Inc.
— Dublin, OH
FDA device recall Z-0728-2020 was initiated by Cardinal Health Inc. on November 14, 2019 and is designated Class II. Reason for recall: Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe". The recall status is ongoing. Affected quantity: 267,247 units total (262,641 OUS, 4,606 US).
Recall Details
- Product Type
- Devices
- Report Date
- January 1, 2020
- Initiation Date
- November 14, 2019
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 267,247 units total (262,641 OUS, 4,606 US)
Product Description
S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) or 7 French (2.3 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The SMART Control stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The stent is equipped with 12 tantalum markers, 6 at each end, to increase the visibility of the stent under fluoroscopic imaging. The SMART Control stent system is designed to be delivered through an appropriately sized catheter sheath introducer.
Reason for Recall
Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".
Distribution Pattern
Worldwide distribution - US Nationwide including Puerto Rico and countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaiian, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Repulic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Maldives, Malta, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saint-Barthelemy, Saudia Arabia, Sinapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela.
Code Information
All Lot Codes. Affected SKUs: N1020ABR, N1020BBR, N1020TBR, N1040ABR, N1040BBR, N1040EBR, N1040TBR, N1060ABR, N1060BBR, N1060EBR, N1060TBR, N1080ABR, N1080BBR, N1080EBR, N1080TBR, N6120BV, N620ABR, N620BBR, N620TBR, N640ABR, N640BBR, N640TBR, N660ABR, N660BBR, N660TBR, N680ABR, N680BBR, N680TBR, N7120BV, N720ABR, N720BBR, N720TBR, N740ABR, N740BBR, N740TBR, N760ABR, N760BBR, N760TBR, N780ABR, N780BBR, N780TBR, N8120BV, N820ABR, N820BBR, N820TBR, N840ABR, N840BBR, N840EBR, N840TBR, N860ABR, N860BBR, N860EBR, N860TBR, N880ABR, N880BBR, N880EBR, N880TBR, N920ABR, N920BBR, N920TBR, N940ABR, N940BBR, N940TBR, N960ABR, N960BBR, N960TBR, N980ABR, N980BBR, N980TBR.