Z-0728-2023 Class II Terminated

Recalled by Icotec Ag — Altstatten Sg

FDA device recall Z-0728-2023 was initiated by Icotec Ag on Nov 3, 2022 and is designated Class II. Reason for recall: During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm… The recall status is terminated (terminated Sep 10, 2026). Affected quantity: 6 devices.

Recall details

Product Type
Devices
Report Date
Dec 28, 2022
Initiation Date
Nov 3, 2022
Termination Date
Sep 10, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 devices

Product description

VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702

Reason for recall

During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.

Distribution pattern

US Nationwide distribution in the state of GA.

Code information

UDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0728-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Icotec Ag? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0728-2023. Report a data issue: hello@regdatalab.com.