Z-0730-2020 Class II Ongoing

Recalled by Cardinal Health Inc. — Dublin, OH

FDA device recall Z-0730-2020 was initiated by Cardinal Health Inc. on Nov 14, 2019 and is designated Class II. Reason for recall: Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe". The recall status is ongoing. Affected quantity: 267,247 units total (262,641 OUS, 4,606 US).

Recall details

Product Type
Devices
Report Date
Jan 1, 2020
Initiation Date
Nov 14, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
267,247 units total (262,641 OUS, 4,606 US)

Product description

Cordis S.M.A.R.T. CONTROL /S.M.A.R.T. Vascular Stent System. The S.M.A.R.T. CONTROL and S.M.A.R.T. Vascular Stent Systems consist of a self-expanding stent made of Nitinol (nickel-titanium alloy) material that is pre-mounted on an over-the-wire delivery system. The stents contain tantalum markers at each end. The S.M.A.R.T. CONTROL and S.M.A.R.T. Vascular Stents are designed to open to a preprogrammed diameter at body temperature. There are two delivery system configurations S.M.A.R.T. CONTROL and S.M.A.R.T. The stent is contained within the outer sheath of the delivery system. Once the distal end of the delivery system reaches the treatment site, the outer sheath of the delivery system is retracted to expose the stent and start its self-expansion. For the S.M.A.R.T. CONTROL Stent System, sheath retraction is achieved by grasping the handle in a fixed position with the tuning dial held between the thumb and index fingers, and rotating the tuning dial in a clockwise direction until the distal end of the stent is visibly apposed to the vessel wall. For the S.M.A.R.T. Stent System, sheath retraction is achieved by grasping the inner shaft in a fixed position and moving the outer sheath proximally relative t the inner shaft.

Reason for recall

Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".

Distribution pattern

Worldwide distribution - US Nationwide including Puerto Rico and countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaiian, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Repulic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Maldives, Malta, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saint-Barthelemy, Saudia Arabia, Sinapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela.

Code information

All Lot Codes. Affected SKUs: C06020MB, C06020MV, C06020SB, C06020SV, C06030MB, C06030MV, C06030SB, C06030SV, C06040MB, C06040MV, C06040SB, C06040SV, C06060MB, C06060MV, C06060SB, C06060SV, C06080MB, C06080MV, C06080SB, C06080SV, C06100MB, C06100MV, C06100SB, C06100SV, C07020MB, C07020MV, C07020SB, C07020SV, C07030MB, C07030MV, C07030SB, C07030SV, C07040MB, C07040MV, C07040SB, C07040SV, C07060MB, C07060MV, C07060SB, C07060SV, C07080MB, C07080MV, C07080SB, C07080SV, C07100MB, C07100MV, C07100SB, C07100SV, C08020MB, C08020MV, C08020SB, C08020SV, C08030MB, C08030MV, C08030SB, C08030SV, C08040MB, C08040MV, C08040SB, C08040SV, C08060MB, C08060MV, C08060SB, C08060SV, C08080MB, C08080MV, C08080SB, C08080SV, C08100MB, C08100MV, C08100SB, C08100SV, C09020MB, C09020MV, C09020SB, C09020SV, C09030MB, C09030MV, C09030SB, C09030SV, C09040MB, C09040MV, C09040SB, C09040SV, C09060MB, C09060MV, C09060SB, C09060SV, C09080MB, C09080MV, C09080SB, C09080SV, C09100MB, C09100MV, C09100SB, C09100SV, C10020MB, C10020MV, C10020SB, C10020SV, C10030MB, C10030MV, C10030SB, C10030SV, C10040MB, C10040MV, C10040SB, C10040SV, C10060MB, C10060MV, C10060SB, C10060SV, C10080MB, C10080MV, C10080SB, C10080SV, C10100MB, C10100MV, C10100SB, C10100SV, C12020MB, C12020MV, C12020SB, C12020SV, C12030MB, C12030MV, C12030SB, C12030SV, C12040MB, C12040MV, C12040SB, C12040SV, C12060MB, C12060MV, C12060SB, C12060SV, C12080MB, C12080MV, C12080SB, C12080SV, C14030MB, C14030MV, C14030SB, C14030SV, C14040MB, C14040MV, C14040SB, C14040SV, C14060MB, C14060MV, C14060SB, C14060SV, C14080MB, C14080MV, C14080SB, C14080SV.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0730-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.