Z-0732-2013 Class II Terminated

Recalled by ConMed Corporation — Utica, NY

FDA device recall Z-0732-2013 was initiated by ConMed Corporation on Dec 27, 2012 and is designated Class II. Reason for recall: ConMed received complaints of some devices which had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products. In no… The recall status is terminated (terminated May 14, 2018). Affected quantity: 313,190 (total for all devices).

Recall details

Product Type
Devices
Report Date
Jan 30, 2013
Initiation Date
Dec 27, 2012
Termination Date
May 14, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
313,190 (total for all devices)

Product description

UltraClean(R) Accessory Electrode 1 inch Coated Needle with Extended Insulation, REF/Catalog No. 139105EXT, Single Use, Sterile, Rx ONLY -- CONMED CORPORATION Electrosurgical, Cutting and Coagulation and Accessories.

Reason for recall

ConMed received complaints of some devices which had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products. In no instance has it been reported to ConMed that a compromise in the sterile barrier has resulted in illness or injury. The compromised seals were discovered prior to use.

Distribution pattern

Worldwide Distribution - USA (nationwide) Puerto Rico and internationally to Belgium, Colombia, Czech Republic, Dubai, Guatemala, Hong Kong, India, Israel, Italy, Japan, Kenya, Kuwait, Malaysia, Mexico, New Zealand, Norway, Oman, Pakistan, Portugal, Saudi Arabia, Thailand, and Uruguay.

Code information

Beginning Lot Code 120628X, Ending Lot Code 120917X. [Lot codes on boxes and packaging contain a lot code in the following format: 120917X where 12 represents the year (2012), 09 represents the month (September), 17 represents the day of the month, and X represents the manufacturing shift code (1, 2, 3, 4,5,6 or 9)].

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0732-2013
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for ConMed Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0732-2013. Report a data issue: hello@regdatalab.com.