Z-0732-2017 Class II Terminated
Recalled by GF Health Products, Inc
— Fond Du Lac, WI
FDA device recall Z-0732-2017 was initiated by GF Health Products, Inc on Sep 30, 2016 and is designated Class II. Reason for recall: GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America "DUOMAT 3" Head/Foot motor used in the manufacture of our Lumex… The recall status is terminated (terminated Apr 29, 2019). Affected quantity: 3,547 (3063 beds, 484 motors).
Recall details
- Product Type
- Devices
- Report Date
- Dec 21, 2016
- Initiation Date
- Sep 30, 2016
- Termination Date
- Apr 29, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3,547 (3063 beds, 484 motors)
Product description
Lumex, Patriot Full Electric Homecare Bed, & Patriot LX Full Electric Homecare Bed containing the Okin Dumat 3 Head/Foot Motors. Description Part Number Motor and Pendant Patriot Full Electric 120V 690-3001-943 / 690-7001-943 Boxed Full Electric 220V Motor 690-3220-943 Motor and Pendant Patriot LX Full Electric 120V 690-8004-912
Reason for recall
GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America "DUOMAT 3" Head/Foot motor used in the manufacture of our Lumex Patriot Full Electric Bed Model US0458/US0468 and Patriot LX Full Electric Bed Model US6000. The failure mode is directly related to contaminated material in the plastic housing used to produce the DUOMAT 3 motor. Contaminated motor housings may crack resulting in a motor that will no longer actuate the head section per the intended design. If the head section were to fall it is unlikely but possible an injury could occur. No injuries have been reported.
Distribution pattern
Worldwide Distribution - USA ( nationwide) Distribution including Puerto Rico and to the countries of : Bahamas, Canada, Chile, El Salvador and Santo Domingo.
Code information
BED SERIAL NUMBER RANGE US0458 / US0468 Patriot Full Electric 120V 697142145108 through 697142147927 US0458 / US0468 Patriot Full Electric 220V 697144140132 through 697144140172 US6000 Patriot LX Full Electric 120V 69204141262 through 69204141737 SERVICE MOTOR KITS WITH SHIP DATES OF MAY 4, 2015 THROUGH JUNE 30, 2016 Description Part Number Motor and Pendant Patriot Full Electric 120V 690-3001-943 / 690-7001-943 Boxed Full Electric 220V Motor 690-3220-943 Motor and Pendant Patriot LX Full Electric 120V 690-8004-912
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0732-2017
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for GF Health Products, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for GF Health Products, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about GF Health Products, Inc →
Use it in ChatGPT
Get a free API key
How the connector works