Z-0736-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 24, 2024
- Initiation Date
- November 16, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5,085,120 units
Product Description
PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547
Reason for Recall
An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.
Distribution Pattern
US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX,UT, VA, WA, WI, WV.
Code Information
UDI-DI: 30382903065470; Lot Numbers: 3207548 3207555 3214826 3226386 3214828 3226388 3250302 3250303 3250304