Z-0736-2024 Class II Ongoing

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

FDA device recall Z-0736-2024 was initiated by Becton Dickinson & Company on Nov 16, 2023 and is designated Class II. Reason for recall: An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes… The recall status is ongoing. Affected quantity: 5,085,120 units.

Recall details

Product Type
Devices
Report Date
Jan 24, 2024
Initiation Date
Nov 16, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,085,120 units

Product description

PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547

Reason for recall

An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.

Distribution pattern

US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX,UT, VA, WA, WI, WV.

Code information

UDI-DI: 30382903065470; Lot Numbers: 3207548 3207555 3214826 3226386 3214828 3226388 3250302 3250303 3250304

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0736-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0736-2024. Report a data issue: hello@regdatalab.com.