Z-0739-2024 Class II Ongoing

Recalled by Smiths Medical Asd Inc — Dublin, OH

Recall Details

Product Type
Devices
Report Date
January 24, 2024
Initiation Date
November 10, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
31,685 total

Product Description

Medex LOGICAL CATH LAB KIT, List Number M20754

Reason for Recall

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, UAE.

Code Information

UDI-DI: 10351688510686; Lot Number: 4371054