Z-0745-2013 Class III Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

FDA device recall Z-0745-2013 was initiated by Siemens Healthcare Diagnostics, Inc. on Dec 20, 2012 and is designated Class III. Reason for recall: Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots… The recall status is terminated (terminated Jan 8, 2014). Affected quantity: 5999.

Recall details

Product Type
Devices
Report Date
Feb 6, 2013
Initiation Date
Dec 20, 2012
Termination Date
Jan 8, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5999

Product description

Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010. The Dimension(R) IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the Iron method for the Dimension(R) clinical chemistry system.

Reason for recall

Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.

Distribution pattern

Nationwide Distribution.

Code information

Siemens material number 10445010 -- and lot numbers 2DN060, exp 2013-10-01; 2FN063, exp 2013-12-01; 2GN068, exp 2014-01-01; and 2JN073, exp 2014-03-01.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0745-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0745-2013. Report a data issue: hello@regdatalab.com.