Z-0745-2013 Class III Terminated
FDA device recall Z-0745-2013 was initiated by Siemens Healthcare Diagnostics, Inc. on Dec 20, 2012 and is designated Class III. Reason for recall: Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots… The recall status is terminated (terminated Jan 8, 2014). Affected quantity: 5999.
Recall details
- Product Type
- Devices
- Report Date
- Feb 6, 2013
- Initiation Date
- Dec 20, 2012
- Termination Date
- Jan 8, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5999
Product description
Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010. The Dimension(R) IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the Iron method for the Dimension(R) clinical chemistry system.
Reason for recall
Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.
Distribution pattern
Nationwide Distribution.
Code information
Siemens material number 10445010 -- and lot numbers 2DN060, exp 2013-10-01; 2FN063, exp 2013-12-01; 2GN068, exp 2014-01-01; and 2JN073, exp 2014-03-01.