Z-0745-2017 Class II Ongoing
FDA device recall Z-0745-2017 was initiated by Becton Dickinson & Company on Aug 8, 2016 and is designated Class II. Reason for recall: BD has received multiple reports of units associated with the 23 G x .75 in. BD Vacutainer¿ Push Button Blood Collection Set with 12 in. Tubing whose lots are exhibiting separation of front and… The recall status is ongoing. Affected quantity: 1,788,150 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 21, 2016
- Initiation Date
- Aug 8, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,788,150 units
Product description
BD 23 G x .75 in. BD Vacutainer¿ Push Button Blood Collection Set with 12 in. Tubing Catalog # 367342 Intended for venipuncture to obtain blood specimens from patients.
Reason for recall
BD has received multiple reports of units associated with the 23 G x .75 in. BD Vacutainer¿ Push Button Blood Collection Set with 12 in. Tubing whose lots are exhibiting separation of front and rear barrels upon activation of the safety feature, which retracts the needle.
Distribution pattern
Worldwide distribution- US (nationwide) and countries of: Canada, Belgium, Brazil.
Code information
5355531 5355533 5356681 5356682 6006629 6006630 6006631 6006633 6006634 6006650
Source
- Dataset