Z-0749-2024 Class II Ongoing

Recalled by Waldemar Link GmbH & Co. KG (Mfg Site) — Norderstedt, N/A

Recall Details

Product Type
Devices
Report Date
January 24, 2024
Initiation Date
November 30, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product Description

Link Patella Glide Resection Guide- For the implantation of the Endo-Model Knee System as part of a total knee arthroplasty procedure. Product Code: 15-2530/05

Reason for Recall

Mislabeled: Angle and height of the device marked as L(large) corresponds to the small/medium dimensions due to a product mix-up. This may result in too little bone material resected and not fit correctly leading to extended surgery time

Distribution Pattern

US Nationwide distribution in the state of NH.

Code Information

UDI-DI :04026575245253 Lot / Serial Number: C306165