Z-0752-2013 Class II Terminated

Recalled by Extremity Medical LLC — Parsippany, NJ

FDA device recall Z-0752-2013 was initiated by Extremity Medical LLC on Mar 22, 2012 and is designated Class II. Reason for recall: Lag screw measurement is short (approximately 5 mm) when using current screw measuring device (p/n 101-00010) per the IO FIX Surgical Technique Rev C. which can lead to possible inappropriate implant… The recall status is terminated (terminated Jun 13, 2013). Affected quantity: 93 units.

Recall details

Product Type
Devices
Report Date
Feb 6, 2013
Initiation Date
Mar 22, 2012
Termination Date
Jun 13, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
93 units

Product description

Extremity Medical, IO FiX System Accessory. Catalog Number 101-00010. This screw measuring device is an accessory to the IO FiX Screw and Washer System that is used to measure the length of the screw by placing the depth gauge over the guidewire and down to the bone. Orthopedic device.

Reason for recall

Lag screw measurement is short (approximately 5 mm) when using current screw measuring device (p/n 101-00010) per the IO FIX Surgical Technique Rev C. which can lead to possible inappropriate implant size selection (e.g. shorter screw size selected).

Distribution pattern

Worldwide Distribution -- Nationwide Distribution including the states of AR, AZ, CA, FL, GA, IL, IN, MD, MT, NC, NE, NH, NJ, NY, OH, PA, TN, TX, WI and WA.

Code information

Catalog Number 101-00010 Lot number AP3V12 AP4E01 AP4N04 AP5E25 AP6A44 AP6E42 AP6L14 AP6M17 AP6T08

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0752-2013
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Extremity Medical LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works