Z-0752-2015 Class II Terminated

Recalled by Stryker Endoscopy — San Jose, CA

FDA device recall Z-0752-2015 was initiated by Stryker Endoscopy on November 17, 2014 and is designated Class II. Reason for recall: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle steri… The recall status is terminated (terminated July 17, 2015). Affected quantity: 180,573 units total.

Recall Details

Product Type
Devices
Report Date
December 24, 2014
Initiation Date
November 17, 2014
Termination Date
July 17, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
180,573 units total

Product Description

PKG, 5MM X 45CM INSERT, DISSECTING FORCEPS, P/N 0250080762 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason for Recall

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Code Information

444728 940354 1045171 1340995 1341513 1441558 1441959