Z-0764-2025 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 22, 2025
- Initiation Date
- December 16, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13,600 units
Product Description
Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223
Reason for Recall
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
Distribution Pattern
US Nationwide Distribution
Code Information
1)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000128741; 2)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000134505; 3)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000133893; 4)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000135769; 5)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000139596; 6)949000222UDI-D:10197344018550(each)30197344018554(case), Lot Number:0000133018; 7)949000224UDI-D:10197344018642(each)30197344018646(case), Lot Number:0000136573; 8)949000221UDI-D:10197344018567(each)30197344018561(case), Lot Number:0000133036; 9)949000223UDI-D:10197344018543(each)30197344018547(case), Lot Number:0000133031;