Z-0765-2016 Class II Terminated
FDA device recall Z-0765-2016 was initiated by Philips Electronics North America Corporation on Nov 3, 2015 and is designated Class II. Reason for recall: Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and… The recall status is terminated (terminated Aug 19, 2016). Affected quantity: 2 subject to correction and removal.
Recall details
- Product Type
- Devices
- Report Date
- Feb 17, 2016
- Initiation Date
- Nov 3, 2015
- Termination Date
- Aug 19, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 subject to correction and removal
Product description
Allura Xper FD OR Table, Ceiling version only; 722035; imaging applications.
Reason for recall
Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).
Distribution pattern
Worldwide Distribution. US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia.
Code information
Ceiling version only.
Source