Z-0766-2013 Class II Terminated

Recalled by Tobii Assistive Technology, Inc. — Dedham, MA

FDA device recall Z-0766-2013 was initiated by Tobii Assistive Technology, Inc. on Jan 14, 2013 and is designated Class II. Reason for recall: Tobii ATI customers P10 external battery pack, which contains sealed lead-acid batteries, malfunctioned while recharging and produced smoke with a chemical smell. The recall status is terminated (terminated Jul 6, 2016). Affected quantity: 483 units.

Recall details

Product Type
Devices
Report Date
Feb 13, 2013
Initiation Date
Jan 14, 2013
Termination Date
Jul 6, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
483 units

Product description

P10 Tobii battery box (product number 510410) Product Usage - Battery pack for MyTobii P10 Device. MyTobii P10 is a portable eye-controlled communication device featuring the Tobii Eye Control system.

Reason for recall

Tobii ATI customers P10 external battery pack, which contains sealed lead-acid batteries, malfunctioned while recharging and produced smoke with a chemical smell.

Distribution pattern

US Nationwide Distribution

Code information

Product Code 510410. All units.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0766-2013
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Tobii Assistive Technology, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0766-2013. Report a data issue: hello@regdatalab.com.